Test Reporting - Acid Fast Stain Result

What is the expected turnaround time for result reporting of acid-fast stains for laboratories that refer acid-fast stains to a reference laboratory?

Any examples are for illustrative purposes only.

Joint Commission Standards require that laboratories report the results of acid-fast stains, both positive and negative, within 24 hours of specimen receipt. The intent is to require the TESTING laboratory to report results of acid-fast stains within 24 hours of specimen receipt. For laboratories that perform acid-fast stains, the expected turnaround time for result reporting is 24 hours.

For laboratories that refer acid-fast stains to reference laboratories, The Joint Commission recommends that the reference laboratories be informed of the 24 hour turnaround time expectation for result reporting by the testing laboratory. In addition, laboratories are encouraged to establish a 24 hour turnaround time for the processing and transport of acid-fast staining specimens to the reference laboratories.
Manual: Laboratory
Chapter: Quality System Assessment for Nonwaived Testing QSA
First published date: April 11, 2016 This Standards FAQ was first published on this date.
This page was last updated on November 01, 2021
Was this response helpful?

If no, please comment on how we could improve this response.

If you have additional standards-related questions regarding this topic, please use the Standards Online Submission Form

Get Extra Help with Books and E-books

Browse our gallery of books and e-books to find trusted prep and readiness resources, practical checklists and toolkits, and resources on specialized health care topics.

Stay Informed with Online Education

Online education is the most convenient and cost-effective way to educate your staff and minimize expenses without the need to leave your organization.

Reach Your Safety Goals with eProducts

Assessing and sustaining compliance with accreditation standards or CMS CoPs can be a challenge, but it doesn’t have to be. Rely on our proven software solutions developed by our team of industry experts.

Can't Find What You're Looking For?

If you do not find an answer to your question, please contact the Standards Interpretation Group (SIG).

Note: To provide adequate support to those organizations that are either accredited/certified or seeking accreditation/certification, we will only answer those questions submitted by those organizations seeking accreditation/certification or currently accredited/certified by the Joint Commission. The Joint Commission no longer answers questions submitted by students or vendors. Thank you for your understanding.