to main content Instrument Reprocessing - Immediate Use Steam Sterilization (IUSS) | Hospital and Hospital Clinics | Infection Prevention and Control IC | The Joint Commission
Instrument Reprocessing - Immediate Use Steam Sterilization (IUSS)

What are important considerations associated with Immediate-Use Steam Sterilization?

Any examples are for illustrative purposes only.

Immediate-Use Steam Sterilization (IUSS), formerly termed "flash" sterilization, is described as "the shortest possible time from the item being removed from the sterilizer to the aseptic transfer onto the sterile field". IUSS items are not intended to be stored for future use.

Considerations for IUSS

  • Review and adhere to manufacturer instructions for use (IFU) to determine if the device or instrument may be reprocessed via IUSS. If so, follow the IFUs regarding cycle type, temperature setting, exposure time, and drying times.
  • IUSS does not imply that reprocessing steps, such as appropriate cleaning and transport, may be omitted. ,.
  • Items are to be reprocessed in approved/validated containers/trays suitable for IUSS.
  • IUSS should not be used for mere convenience, or due to limited instruments or equipment for the number of cases/procedures performed.

Evidence-based guidelines should be adopted to minimize the use of IUSS.  Scenarios when IUSS may be appropriate include:

  • When a specific instrument is needed for an emergency procedure.
  • When a non-replaceable instrument has been contaminated and needs to be replaced to the sterile field immediately.
  • When an item has dropped on the floor and is needed to continue a surgical procedure.
Identifying Gaps and Risk reduction strategies to consider
  • When 'loaner' trays or instruments (including those brought in by a provider) are used, establish an agreement with the vendor/provider requiring that delivery occurs sufficiently in advance of scheduled case(s)to allow complete reprocessing of trays by the organization.  Such a requirement is an example of a performance expectation to include in a contract (see LD.04.03.09).
  • Develop policies, procedures, staff orientation and competencies based on evidence-based guidelines. 
  • Regular rounding by Leadership experienced in sterilization practices to all areas where instruments are used and reprocessed is critical. Such rounds should include:
Review of manufacturer's IFUs for both the device used for IUSS as well as the equipment/instruments being reprocessed to ensure compliance with their guidelines. For example:
  • Allowing sufficient time to actually observe reprocessing activities, including a review of any documentation requirements.
  • Soliciting questions/concerns from staff responsible for performing IUSS and implement plans to reduce/eliminate concerns.
  • Ensure there is a defined, evidence-based process in place for the premature release of items, to include documentation of IUSS.
  • When limited resources are identified, work with Leadership to develop a plan to ensure sufficient resources are available to support the delivery of safe, quality of care. 
  • Ensure that sterilization practices, in all locations, have been fully incorporated into the organization's Quality Assessment Performance Improvement (QAPI) activities.
  • Evaluate the IUSS process in all locations that it is being performed.  Surveyors evaluate compliance based on the evidence-based guidelines, policies, procedures, practices and competencies adopted by the organization.
Conducting a risk assessment is a helpful way of identifying risks and gaps in compliance with evidence-based guidelines and product manufacturer's instructions for use.  A proactive risk assessment examines a process in detail including sequencing of events, actual and potential risks, and failure or points of vulnerability and that prioritizes, through a logical process, areas for improvement based on the actual or potential impact (that is, criticality) of care, treatment, or services provided. The introductory section of the Leadership (LD) chapter provides an example of a pro-active risk assessment model that an organization may use.  However, this specific approach is not mandated as there are other risk assessment tools available that may better meet the needs of the organization.

Resources

  • The Association for Professionals in Infection Control and Epidemiology (APIC)
  • The Association of periOperative Registered Nurses (AORN)
  • The Association for the Advancement of Medical Instrumentation (AAMI).
  • CDC Guideline for Disinfection and Sterilization in Healthcare Facilities. 2008.
Manual: Hospital and Hospital Clinics
Chapter: Infection Prevention and Control IC
First published date: May 12, 2017 This Standards FAQ was first published on this date.
This page was last updated on November 29, 2021 with update notes of: Editorial changes only Types of changes and an explanation of change type: Editorial changes only: Format changes only. No changes to content. | Review only, FAQ is current: Periodic review completed, no changes to content. | Reflects new or updated requirements: Changes represent new or revised requirements.
Was this response helpful?

If no, please comment on how we could improve this response.

If you have additional standards-related questions regarding this topic, please use the Standards Online Submission Form

Get Extra Help with Books and E-books

Browse our gallery of books and e-books to find trusted prep and readiness resources, practical checklists and toolkits, and resources on specialized health care topics.

Stay Informed with Online Education

Online education is the most convenient and cost-effective way to educate your staff and minimize expenses without the need to leave your organization.

Reach Your Safety Goals with eProducts

Assessing and sustaining compliance with accreditation standards or CMS CoPs can be a challenge, but it doesn’t have to be. Rely on our proven software solutions developed by our team of industry experts.

Can't Find What You're Looking For?

If you do not find an answer to your question, please contact the Standards Interpretation Group (SIG).

Note: To provide adequate support to those organizations that are either accredited/certified or seeking accreditation/certification, we will only answer those questions submitted by those organizations seeking accreditation/certification or currently accredited/certified by the Joint Commission. The Joint Commission no longer answers questions submitted by students or vendors. Thank you for your understanding.